5322A
Calibration Manual
54
Table 12. RCD Trip Current Adjustment Points
6. The Product shows the value entered in the last adjustment.
7. From the 5522A AUX output, set the value to 30 mA at 55 Hz. Push the ON button on the
5522A.
8. The Product reads approximately 30 mA. Use the cursor buttons or rotary knob to set the
measured value to 30.000 mA.
9. Push the
Write
softkey to save the new value for the 30 mA current range. To leave the
original value, push the
Exit
softkey. The Product ignores the measured current and returns
to list of current ranges.
10. Repeat these steps for the next two ranges.
11. Highlight
Line Voltage
and push the
Select
softkey. The Product shows line voltage and
frequency.
12. Connect the 8588A to the L and N mains terminals, see
13. Set the range to 1000 V ac. The 5522A measures the power line voltage.
Nominal Current
Range
Position
Adjustment Point
Frequency
Required
Accuracy of
Adjustment
30 mA
I01
25 mA
55 Hz
0.2 % + 50
μ
A
300 mA
I02
250 mA
55 Hz
0.2 %
3 000 mA
I03
2 500 mA
55 Hz
0.2 %
30 mA
I04
25 mA
55 Hz
0.2 % + 50
μ
A
150 mA
I05
125 mA
55 Hz
0.2 %