Table 2: Internal alarms
When multiple alarms occur, the error code shown is the sum of the error codes as given below. For
example 0048 is a combination of error code 0016 and 0032.
Alarm
Explanation
0001
Display error
0002
Data storage error
0004
Initialization error
0016
IO configuration error base
0032
IO configuration error interface
0064
Analog input error
0128
Automatic K-factor conversion error
0256
PT100 ADC error (type TP only)
0512
Total S-factor calculation error
1024
TPC factor calculation error
2048
Temperature sensor out of range error