Chapter 4: Maintenance
Autosampler Module Maintenance
CyTOF XT User Guide
97
9
Click
Move To Home
to raise the probe line out of the tube and back to is default
position.
Replace A Bottle Line
NOTE
We recommend removing 1 line at a time to avoid mistakes.
1
Remove the nut of the damaged bottle line from the bottle.
2
Remove the bottle line from the tube guides and the hole on the Autosampler top plate.
3
Gently remove the bottle line from the groove on the syringe guard.
4
Remove the other end of the bottle line from its place on the pump and valve unit.
5
Connect the new bottle line to its position on the pump and valve unit.
6
Gently press the tube into the groove on the syringe guard.
7
Feed the bottle line through the tube guides and then through the guide hole on the
Autosampler top plate.
8
Connect the new bottle line to the bottle cap.
9
Repeat Steps 1
–
8 for additional bottle lines one at a time if necessary.