![Fluidigm CyTOF XT User Manual Download Page 93](http://html1.mh-extra.com/html/fluidigm/cytof-xt/cytof-xt_user-manual_2308832093.webp)
Chapter 4: Maintenance
Autosampler Module Maintenance
CyTOF XT User Guide
93
2
Pull down on the quick-release connection to unlock the probe line from its position on
the stir assembly.
3
Remove the probe line from the stir assembly.
4
Thread the new probe line through the stir assembly and lock the probe line in place
using the quick release. You should hear an audible click.
5
Insert the other end of the probe line in the PROBE (gold) position of the selector valve
on the pump and valve unit and tighten by turning clockwise until finger-tight, taking
care to not leave an air gap in between the end of the line and the valve port.
6
Recalibrate the height of the new probe line as instructed below.
Recalibrate the Probe Line
1
Place an empty tube in the carousel. Note the location number.
2
In the CyTOF Software, on the
Instrument Control
tab, click
Autosampler camera
. The
tube below the probe line in the carousel is now visible.
Summary of Contents for CyTOF XT
Page 1: ...FLDM 00254 Rev 03...