Chapter 4: Maintenance
Autosampler Module Maintenance
92
CyTOF XT User Guide
4
Slide the sample loop agitator motor into the new sample loop and reinsert the sample
loop agitator connector onto the pump and valve unit.
Replace the Probe Line
1
Turn the probe line counterclockwise to loosen the fitting and then remove the probe
line in the PROBE position on the selector valve of the pump and valve unit.