Chapter 4: Maintenance
Clean the Vacuum Interface Cones
70
CyTOF XT User Guide
10
Remove the skimmer-reducer assembly from the cone removal tool.
NOTE
Place the tip up to avoid damage to the cone orifice.
Clean Cones
1
Stack the cones as follows in the cone cleaning container:
IMPORTANT
Add 10% Citranox as each part is placed in the container to ensure that all
parts contact the cleaning solution.
a
Stack the bottom adapter first.
b
Place the sampler cone tip face down on the bottom adapter.
Summary of Contents for CyTOF XT