Chapter 4: Maintenance
Nebulizer Assembly Maintenance
CyTOF XT User Guide
111
Replace the Nebulizer Gas Line
1
Pull to disconnect the quick connect nebulizer gas line from the nebulizer assembly.
2
Remove the nebulizer gas from the CyTOF XT instrument using an adjustable wrench.
3
Replace the line with a new nebulizer gas line and tighten it using an adjustable wrench.
4
Push to connect the quick connect end of the nebulizer gas line to the nebulizer
assembly.
NOTE
An audible click is heard when it is connected.
Clean the Nebulizer Mount
1
When cleaning is required, remove the nebulizer holder from the nebulizer mount as
outlined in the