10.2.3.5
Obj par
10439303;a2
Figure 10.17
Object Parameters
dialog box
You use this command to set the object parameters
Emissivity
,
Distance
,
T Reflected
,
T Atmosphere
,
Rel humidity
,
External optics
,
Optics transmission
, and
Optics
temperature
. The parameters are selected by moving the joystick up/down and set
by moving the joystick left/right. These parameters settings will be used by all mea-
surement functions that have not been set locally.
➲
Please note the following:
■
The emissivity file can be stored at root level or at directory level. However, the
camera software prioritizes files that are stored at directory level and the directory
has to be selected in order to store the emissivity file in the camera memory. If the
camera software does not find an emissivity file at directory level, it searches for
similar files at root level and saves those instead.
■
If you enter an emissivity value less than 0.30 the emissivity box will begin flashing
to remind you that this value is unusually low.
■
The transmission factor is applied to the signal and not to the temperature
For more information about object parameters, see section 18 – Thermographic
measurement techniques on page 125.
10.2.3.6
Deactivate local par.
Point to
Deactivate local par.
and press the joystick to delete all locally set parameters.
Locally set parameters are the parameters you set in
e.g.
the
Spot settings
dialog
box.
10
74
Publ. No. 1557978 Rev. a155 – ENGLISH (EN) – February 6, 2006
10 – Camera program
Summary of Contents for ThermaCam P25
Page 2: ......
Page 4: ......
Page 6: ......
Page 7: ...ThermaCAM P25 User s manual Publ No 1557978 Rev a155 ENGLISH EN February 6 2006...
Page 66: ...INTENTIONALLY LEFT BLANK 8 52 Publ No 1557978 Rev a155 ENGLISH EN February 6 2006 8 Tutorials...
Page 191: ......