Action
Step
Measure the apparent temperature of the aluminum foil and write it down.
10727003;a2
Figure 16.4 Measuring the apparent temperature of the aluminum foil
5
16.2.1.2
Step 2: Determining the emissivity
Action
Step
Select a place to put the sample.
1
Determine and set reflected apparent temperature according to the previous pro-
cedure.
2
Put a piece of electrical tape with known high emissivity on the sample.
3
Heat the sample at least 20 K above room temperature. Heating must be reasonably
even.
4
Focus and auto-adjust the camera, and freeze the image.
5
Adjust
Level
and
Span
for best image brightness and contrast.
6
Set emissivity to that of the tape (usually 0.97).
7
Measure the temperature of the tape using one of the following measurement
functions:
■
Isotherm
(helps you to determine both the temperature and how evenly you
have heated the sample)
■
Spot
(simpler)
■
Box Avg
(good for surfaces with varying emissivity).
8
Write down the temperature.
9
Move your measurement function to the sample surface.
10
Change the emissivity setting until you read the same temperature as your previous
measurement.
11
16
112
Publ. No. 1558407 Rev. a155 – ENGLISH (EN) – February 6, 2006
16 – Thermographic measurement techniques
Summary of Contents for ThermaCAM E320
Page 2: ......
Page 4: ......
Page 6: ......
Page 7: ...ThermaCAM E320 User s manual Publ No 1558407 Rev a155 ENGLISH EN February 6 2006...
Page 70: ...INTENTIONALLY LEFT BLANK 8 56 Publ No 1558407 Rev a155 ENGLISH EN February 6 2006 8 Tutorials...
Page 169: ......