Thermographic measurement techniques
41.3
Reflected apparent temperature
This parameter is used to compensate for the radiation reflected in the object. If the
emissivity is low and the object temperature relatively far from that of the reflected it will
be important to set and compensate for the reflected apparent temperature correctly.
41.4
Distance
The distance is the distance between the object and the front lens of the camera. This
parameter is used to compensate for the following two facts:
• That radiation from the target is absorbed by the atmosphere between the object and
the camera.
• That radiation from the atmosphere itself is detected by the camera.
41.5
Relative humidity
The camera can also compensate for the fact that the transmittance is also dependent
on the relative humidity of the atmosphere. To do this set the relative humidity to the cor-
rect value. For short distances and normal humidity the relative humidity can normally be
left at a default value of 50%.
41.6
Other parameters
In addition, some cameras and analysis programs from FLIR Systems allow you to com-
pensate for the following parameters:
• Atmospheric temperature –
i.e.
the temperature of the atmosphere between the cam-
era and the target
• External optics temperature –
i.e.
the temperature of any external lenses or windows
used in front of the camera
• External optics transmittance –
i.e.
the transmission of any external lenses or windows
used in front of the camera
#T810169; r. AN/42241/42268; en-US
121
Summary of Contents for GFx320
Page 1: ...User s manual FLIR GFx3xx series...
Page 2: ......
Page 3: ......
Page 5: ...User s manual FLIR GFx3xx series T810169 r AN 42241 42268 en US v...
Page 6: ......
Page 92: ...Mechanical drawings 25 See next page T810169 r AN 42241 42268 en US 80...
Page 95: ...EU Declaration of conformity 26 See next page T810169 r AN 42241 42268 en US 83...
Page 96: ......
Page 97: ...MET Compliance Data Report truncated 27 See next page T810169 r AN 42241 42268 en US 85...
Page 100: ...IEC IECEE Intertek Test Report truncated 28 See next page T810169 r AN 42241 42268 en US 88...
Page 102: ......
Page 103: ...IEC IECEE Intertek CB Test Certificate 29 See next page T810169 r AN 42241 42268 en US 91...
Page 104: ......
Page 105: ......
Page 106: ...MET Laboratories Test Certificate truncated 30 See next page T810169 r AN 42241 42268 en US 94...
Page 108: ...MET Laboratories Letter of Certification 31 See next page T810169 r AN 42241 42268 en US 96...
Page 163: ......