FT-10 Fill Technical Manual, Rev. 2.21, July 2020
Page 125 of 131
21.
T
ROUBLE
S
HOOTING
FT-10 FILL filling controller had been designed as a very reliable and virtually error free instrument. However
if there is an error occurs, do not attempt to repair the equipment before understanding what caused the
error. Note the problems you have with your instrument and the error messages shown on the display. Then
try to solve the problem according to the error table given below.
ERROR
CODE
DESCRIBTION
THINGS TO DO
Under
Weight is too low
- Check the load
- Load cell or instrument could be broken.
Over
Over Load
ADC Out
Load exceeds the operation range
- Check the load
- Check the calibration
- Load cell or instrument could be broken.
Err 1
ADC error
- Re-energize indicator
- Call FLINTEC
Err 2
ADC error
- Re-energize indicator
- Call FLINTEC
Err 3
Indicator cannot be calibrating
- Check load cell cable and load then start
calibration again
Err 10
EEPROM error
- Configure the instrument
- EEPROM broken
Err 20
Calibration error
- Calibrate the indicator.
Err 21
Configuration error
- Configure the indicator.
Err 22
Tare, CN, Total weight and the SP in
use error
- Check SP, PT and ID entries.
- Check Tare, CN and Total weight
Err 27
Indicator is not calibrated
- Calibrate the indicator
Err 30
Processor Error
- Call FLINTEC
Err 34
Indicator cannot be calibrating
- Load cell signal is negative, very low or
too high
Err 35
Calibration Error
- Calibration loading is not enough.
- Check test weight loading.
Err 36
Calibration load value entry Error
- Test weight is too small. Increase the test
weight.
Err 37
Scale unstable
- Wait until scale become stable.
- Check grounding wiring.
Err 47
Main pcb info error
- Call FLINTEC
Err 61
Eeprom is not installed or broken
- Call FLINTEC
Err 70
Modbus selection error
- Check data format of other interfaces.
- Other interfaces should not be Modbus.
E XXXX
Hardware error
- Call FLINTEC