Operator's Manual
|
3-4
On / Standby
Figure 2 – On/Standby Button
This button toggles between the following conditions:
Standby: (if attached to external power, the battery is being
charged) -
►
Setting (allows setting of control parameters) -
►
On
(enables ventilation) -
►
-
►
Standby.
Standby: The FLIGHT 50 Ventilator is dormant and ventilation is not
enabled. If attached to external power, the Ext. Power/ Charging
Int. Battery LED is lit green, indicating that the internal battery is
being charged. The On/Standby indicator is not lit.
Setting: Pressing the
On/Standby
button once changes the ventilator
from Standby to Setting condition.
There is approximately a two second delay in going from
Standby to Setting condition. During this time, the FLIGHT
50 Ventilator performs a self-test and will light all displays
on the front panel.
During Setting condition, all adjustable LEDs are lit. This allows the operator
to preset and adjust controls prior to ventilation. The On/Standby indicator
is not lit. The Message Display Window shows
Press On to Vent
, suggesting
to the operator that the
On/Standby
button needs to be pressed if the
FLIGHT 50 Ventilator should start ventilation.
On:
Pressing the
On/Standby
button once more changes the ventilator from
Setting to On. In the On condition, the FLIGHT 50 Ventilator is ventilating
and the On/Standby indicator is lit green.
Pressing the
On/Standby
button twice while in On condition turns the
ventilator from On to Standby.
3.2.1 MODE Control
Figure 3 – Mode Control Button