15 Outputs (TF7-*721**-******-******a-*****-**; a = U, V, D)
PIOX S72*
15.3 Activation of a frequency output as pulse output
151
UMPIOX_S72xV1-9EN, 2022-05-15
15.3
Activation of a frequency output as pulse output
• Select the list item
Frequency
F1
in the menu item
Outputs
.
• Press ENTER.
• Select the list item
Pulse
in the menu item
Source item.
• Press ENTER.
• Select a list item (here:
Pulse +V
).
• Press ENTER.
• Select a list item:
– 0...1 kHz
– 0...5 kHz
– Other range
• Press ENTER.
If
Other range
is selected, enter a value for
Output MAX
.
• Enter the number of pulses per unit.
• Press ENTER.
The unit of measurement will be displayed according to the actual physical quantity.
Terminal assignment
The terminals for the connection of the output are displayed.
By pressing
or
further information is displayed.
• Press
twice.
The max. value of the physical quantity is displayed.
• Press ENTER.
The pulse output emits a frequency resulting from the physical quantity of the fluid passing at the measuring point and the
entered number of pulses per unit.
Options\Channel A\Outputs\Frequency F1
Options\Channel A\Outputs\Frequency F1\Source item\Pulse
Options\Channel A\Outputs\Frequency F1\Source item\Pulse\Pulse +V
Options\...\Frequency F1\Source item\Pulse\F1 Output range
Options\...\Frequency F1\Source item\Pulse\F1 Output range\Pulses per unit
Options\...\Frequency F1\...\Output info
Example
pulse:
Pulse +V
output range:
0...1 kHz
pulses per unit:
100/m3
5 m
3
of the fluid are passing at the measuring point. A frequency of 0.5 kHz is transmitted at the pulse output.
display:
Maximum 36000.00 m3/h