13 Outputs
PIOX R721
13.2 Digital output
71
UMPIOX_F72xV1-3EN, 2021-09-01
• Select
Yes
to test the assignment of the measured value to the output signal. Select
No
to display the next menu item.
• Press ENTER.
• Select a list item as test value.
• Press ENTER.
• Select
Repeat
to repeat the test. Select
Finish
to display the next menu item.
• Press ENTER.
13.2
Digital output
A digital output switches if one of the following switching conditions is met:
– the measured value exceeds or falls below a limit
– the measured value lays within or outside a defined range
– a measurement is not possible
– an event occurs
13.2.1
Configuration of a digital output as binary output
• Select the menu
Outputs
.
• Press ENTER.
• Select
Yes
to change the settings for an assigned output or to assign a new output.
• Select
No
to cancel the assignment and to return to the previous menu item.
• Press ENTER.
Outputs\...\Test mea. range
Outputs\...\Enter test value
status value
event value
Status OK
or
Status error
.
If the external measuring instrument displays the value
(min. output value for
Status error
, max. output for
Status
OK
), the output functions correctly.
Active
or
Passive
.
If the external measuring instrument displays the value
(min. output value for
Passive
, max. output for
Active
),
the output functions correctly.
Outputs\Binary B1\B1 Enable