13 Settings
66
UMFLUXUS_F5WDV1-1EN, 2016-04-08
13.2.1
Pipe Circumference
13.2.2
Measuring Point Number
This display will only be indicated if the transmitter has an RS485 interface.
13.2.3
Transducer Distance
13.2.4
Error Value Delay
The error value delay is the time after which an error value will be sent to an output if no valid measured values are avail-
able.
Select
on
if the pipe circumference is to be entered instead of the pipe diameter in the pro-
gram branch
Parameter
. Press ENTER.
If
on
has been selected for
Pipe Circumfer.
, the outer pipe diameter will nevertheless
be requested in the program branch
Parameter
.
To select the menu item
Pipe Circumfer.
, enter
0
(zero). Press ENTER.
The value displayed in
Pipe Circumfer.
is calculated on the basis of the last displayed
value of the outer pipe diameter.
example: 100 mm
.
π
= 314.2 mm
Enter the pipe circumference. The limits for the pipe circumference are calculated on the
basis of the limits for the outer pipe diameter.
During the next scroll through the program branch
Parameter
, the outer pipe diameter
that corresponds to the entered pipe circumference will be displayed.
example: 180 mm :
π
= 57.3 mm
Note!
The pipe circumference is only edited temporarily. When the transmitter switches back to the display
of the pipe circumference (internal recalculation), slight rounding errors may occur.
Example:
entered pipe circumference: 100 mm
displayed outer pipe diameter: 31.8 mm
When the transmitter switches back to the display of the pipe circumference, 99.9 mm will be dis-
played.
Select
(1234)
if the measuring point is to be identified only by numbers, point and dash.
Select
(
↑↓←
→
)
if the measuring point is to be designated with ASCII characters.
recommended setting:
user
•
user
will be selected if the measuring point is always the same.
•
auto
can be selected if the measuring point changes often.
In the program branch
Measuring
, the recommended transducer distance will be dis-
played in parentheses, followed by the entered transducer distance if the recommended
and the entered transducer distance are not identical.
During transducer positioning in the program branch
Measuring
• only the entered transducer distance will be displayed if
Transd. Distance = user
has been selected and the recommended and the entered transducer distances are iden-
tical
• only the recommended transducer distance will be displayed if
Transd. Distance =
auto
has been selected.
Select
edit
to enter an error value delay. Select
damping
if the damping factor is to be
used as the error value delay.
For further information on the behavior of missing measured values see section 15.1.2 and
15.2.
Pipe Circumfer.
off
>ON<
Outer Diameter
100.0
mm
Pipe Circumfer.
314.2
mm
Pipe Circumfer.
180
mm
Outer Diameter
57.3
mm
Meas.Point No.:
(1234) >(
↑↓←
→
)<
Transd. Distance
auto
>USER<
Transd. Distance?
(50.8) 50.0 mm
Transd. Distance?
50.8
mm
Error-val. delay
damping
>EDIT<
Summary of Contents for FLUXUS WD100
Page 1: ...User Manual UMFLUXUS_F5WDV1 1EN Ultrasonic Flowmeter for Water FLUXUS WD100 FLUXUS WD200...
Page 4: ...FLUXUS WD100 WD200 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Page 8: ...FLUXUS WD100 WD200 Table of Contents 8 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Page 104: ...FLUXUS WD100 WD200 D Declarations of Conformity 104 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Page 105: ......
Page 106: ......