11 Advanced Measuring Functions
52
UMFLUXUS_F5WDV1-1EN, 2016-04-08
11
Advanced Measuring Functions
11.1
Command Execution during Measurement
Commands that can be executed during a measurement are displayed in the upper line. A command begins with the ar-
row
→
. If programmed, a program code has to be entered first (see section 11.8).
Press
until the command is displayed. Press ENTER. The following commands are available:
11.2
Damping Factor
Each displayed measured value is a floating average of all measured values of the last x seconds, with x being the damp-
ing factor. A damping factor of 1 s means that the measured values are not averaged because the measuring rate is ap-
prox 1/s. The default value of 10 s is appropriate for normal flow conditions. Strongly fluctuating values caused by high
flow dynamics require a higher damping factor.
Select the program branch
Output Options
. Press ENTER until the menu item
Damping
is displayed.
Press BRK to return to the main menu.
11.3
Totalizers
Total volume or total mass of the fluid at the measuring point can be determined.
There are two totalizers, one for the positive flow direction, one for the negative flow direction. The unit of measurement
used for totalizing corresponds to the volume or mass unit selected for the physical quantity.
The values of the totalizers can be displayed with up to 11 places, e.g.,
74890046.03
. For the adjustment of the number
of decimal places (max 4) see section 14.6.
Press ENTER while a totalizer is displayed to toggle between the display of the totalizers for the two flow directions.
Tab. 11.1: Commands that can be executed during the measurement
command
explanation
→
Adjust transd.
Select transducer positioning.
If a program code is active, the measurement will be continued 8 s after the last keyboard en-
try.
→
Clear totalizer
All totalizers will be reset to zero.
→
Break measure
stop the measurement and return to the main menu
Enter the damping factor. Press ENTER.
Press key
to scroll through the upper line to the display of the totalizers.
The value of the totalizer will be displayed in the upper line (here: the volume which has
passed through the pipe at the measuring point in the positive flow direction after the acti-
vation of the totalizers).
S=
< >
=54
mm!
32.5
m3
54.5
m3/h
Damping
10
s
Volume flow
54.5
m3/h
32.5
m3
54.5
m3/h
Summary of Contents for FLUXUS WD100
Page 1: ...User Manual UMFLUXUS_F5WDV1 1EN Ultrasonic Flowmeter for Water FLUXUS WD100 FLUXUS WD200...
Page 4: ...FLUXUS WD100 WD200 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Page 8: ...FLUXUS WD100 WD200 Table of Contents 8 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Page 104: ...FLUXUS WD100 WD200 D Declarations of Conformity 104 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Page 105: ......
Page 106: ......