17 Advanced functions
17.2 Diagnosis with the help of the snap function
FLUXUS G831ST-HT
2021-06-01, UMFLUXUS_G831ST-HTV1-0EN
124
17.2
Diagnosis with the help of the snap function
17.2.1
Configuration
By means of the snap function it is possible to store measuring parameters which are useful for the evaluation of
measuring results or diagnostic purposes. To make use of the snap function it has to be configured.
• Select the menu item
Configuration
.
• Press ENTER.
• Select
On
to activate the snap function.
• Press ENTER.
• Select
Yes
to activate the snap ringbuffer.
If the snap ringbuffer is activated, after taking the 51th snap the oldest snaps will be overwritten. If the ringbuffer is
deactivated, up to 50 snaps can be stored.
• Press ENTER.
• Select
Yes
if the auto snap has to be activated.
If the auto snap is activated, snaps are automatically stored during a measurement failure.
• Press ENTER.
• Select
Yes
if an event that releases a snap has been parameterized for the event trigger R1.
• Press ENTER.
17.2.2
Information concerning snaps
• Select the menu item
Snap info
.
• Press ENTER.
The following information is displayed:
Special functions\Snap\Configuration
Special functions\Snap\Configuration\Snap
Special functions\Snap\Configuration\Snap ringbuffer
Special functions\Snap\Configuration\Auto snap
Special functions\Snap\Configuration\Snap on R1
Special functions\Snap\Snap info
display
description
Stored snaps
number of stored snaps
Snaps left
number of snaps that can still be stored
Ringbuffer
snap ringbuffer activated