9 Measurement
9.3 Start of the measurement
FLUXUS G801
2020-06-26, UMFLUXUS_G801V1-0EN
96
9.2.5
Input of the error delay
The error delay is the time after which an error value will be sent to an output if no valid measured values are available.
This display will only be indicated if
edit
is selected in the menu item
Error-val. delay
.
If no error delay is entered, the damping factor will be used.
• Enter a value for the error delay.
• Press ENTER.
9.3
Start of the measurement
• Select the program branch
Measuring
.
• Press ENTER.
If the parameters in the program branch
Parameter
are
not valid or incomplete, the error message
NO DATA!
will
be displayed.
Activation of the channels
The channels for the measurement can be activated and
deactivated.
This display will not be indicated if the transmitter has only one measuring channel.
• Select a channel with
.
• Press
to activate or deactivate the channel.
• Press ENTER.
A deactivated channel will be ignored during the measurement. The parameters entered for this channel will remain
unchanged.
If the data logger or the serial interface is activated, the measuring point number has to be entered:
Input of the measuring point number
• Enter the number of the measuring point.
• Press ENTER.
For the activation of text input, see
Special Funct.\SYSTEM settings\Dialogs/Menus\Meas.Point No.
Output Options\...\I1:Error-val. delay
Measuring
Channel x
Meas.Point No.
Sound Path
Transd. Distance
Signal test
Transd. Distance
Measurement display
see annex A
Measuring\Channel x
the channel is activated
– the channel is deactivated
• the channel cannot be activated
Notice!
A channel cannot be activated if the parameters are not valid, e.g., if the parameters in the program branch
Parameter
of the channel are not complete.
Measuring\Channel\Meas.Point No.