11 Maintenance and cleaning
11.4 Examination of the O-ring
FLUXUS G801
2020-06-26, UMFLUXUS_G801V1-0EN
110
11.4
Examination of the O-ring
The O-rings of the transmitter and the cable glands have to be checked regularly as well as after each opening of the
housing for possible damages. Porous or damaged O-rings have to be replaced. Cable glands with defective O-rings have
to be replaced.
Use original spare parts only. Store the O-ring replacement in a hermetically sealed and dark place.
Housing with increased safety
Defective O-rings have to be replaced.
Danger!
Risk of explosion when using the transmitter in explosive atmospheres
This may result in personal or material damage or dangerous situations.
→ The O-rings form part of the ignition protection and have to ensure tightness of the housing to
guarantee explosion protection.
Important!
O-rings of increased safety housings may only be replaced by qualified FLEXIM personnel.
Fig. 11.1: Transmitter FLUXUS *801
1 – Housing with increased safety
2 – flameproof enclosure housing
3 – O-ring of transmitter
4 – O-ring of cable gland
5 – thread gap (Ex db)