14 Data Logger and Transmission of Data
UMFLUXUS_G8V4-6-1EN, 2018-10-10
99
Storing of the Signal Amplitude
Storing of the Sound Speed of the Fluid
Storing of the Diagnostic Values
Acoustic Signal during the Storing
Per default, an acoustic signal will be emitted every time a measured value is stored or transmitted to a PC or printer. The
signal can be deactivated in
Special Funct.
\
SYSTEM settings
\
Storing
\
Beep on storage
.
14.1.4
Measurement with Activated Data Logger
• Start the measurement.
If
Output Options
\
Store Meas.Data
has been activated and
Special Funct.
\
Ringbuffer
\ is deactivated, this
error message will be displayed as soon as the data logger is full.
Th Storing will be stopped.
14.1.5
Deleting the Measured Values
14.1.6
Available Data Logger Memory
If the data logger is empty and a measurement is started with one physical quantity on one measuring channel without
storing the totalizer, approx. 100 000 measured values can be stored. The available data logger memory can be dis-
played:
If
on
is selected and the data logger is activated, the amplitude of the measured signal will
be stored together with the measured values. Press ENTER.
If
on
is selected and the data logger is activated, the sound speed of the fluid will be stored
together with the measured values. Press ENTER.
If
on
is selected and the data logger is activated, the diagnostic values will be stored to-
gether with the measured values. Press ENTER.
Select
off
to deactivate the acoustic signal,
on
to activate it. Press ENTER.
Enter the measuring point number. Press ENTER.
If arrows are displayed in the lower line on the right, ASCII text can be entered. If digits are
displayed, only digits, point and hyphen can be entered.
For the setting of the input mode see section 16.2.3.
Press ENTER.
The error message will be displayed periodically.
Select
Special Funct.\Delete Meas.Val.
. Press ENTER.
Select
yes
or
no
. Press ENTER.
Select
Special Funct.\Instrum. Inform.
. Press ENTER.
Store Amplitude
off
>ON<
Store c-Medium
off
>ON<
Store diagnostic
off
>ON<
Beep on storage
>ON<
off
A:Meas.Point No.:
xxx (
↑↓←
→
)
DATA MEMORY
OVERFLOW!
Special Funct.
↕
Delete Meas.Val.
Really Delete?
no
>YES<
Special Funct.
↕