18 Outputs
126
UMFLUXUS_G8V4-6-1EN, 2018-10-10
18.1.3
Function Test
The function of the installed output can now be tested. Connect a multimeter to the installed output.
Test of the Analog Outputs
Test of the Binary Outputs
18.2
Error Value Delay
The error value delay is the time interval after which the entered value for the error value is transmitted to the output in
case no valid measured values are available. The error value delay can be entered in the program branch
Output Op-
tions
if this menu item has previously been activated in the program branch
Special Funct.
. If the error value delay
is not entered, the damping factor will be used.
Select a list item for the error output. Press ENTER.
If
Other value
is selected, enter an error value. It has to be within the limits of the output.
Press ENTER.
Note!
The settings will be stored at the end of the dialog.
The terminals for the connection of the output are displayed (here:
1-
and
2+
for the active
current loop).
Press ENTER.
The current output is tested in the display. Enter a test value. It has to be within the output
range. Press ENTER.
If the multimeter displays the entered value, the output functions correctly.
Select
yes
to repeat the test,
no
to return to
SYSTEM settings
. Press ENTER.
Select
Reed-Relay OFF
or
Open collect OFF
in the scroll list
Output Test
to test
the de-energized state of the output. Press ENTER. Measure the resistance at the output.
The value has to be high ohmic.
Select
yes
. Press ENTER.
Select
Reed-Relay ON
or
Open collect. ON
in the scroll list
Output Test
to test the
energized state of the output. Press ENTER. Measure the resistance at the output. The val-
ue has to be low ohmic.
Select
yes
to repeat the test,
no
to return to
SYSTEM settings
. Press ENTER.
Select
Special Funct.\SYSTEM settings\Dialogs/Menus\Error-val. delay
.
Select
damping
if the damping factor is to be used as the error value delay. Select
edit
to
activate the menu item
Error-val. delay
in the program branch
Output Options
.
From now on, the error value delay can be entered in the program branch
Output Op-
tions
.
Error-value
↕
Minimum (4.0mA)
Error-value
3.5
mA
I1 active loop
Terminal:1-,2+
I1:Output Test
4
mA
I1= 4.0 mA
Again? no
>YES<
B1:Output Test
↕
Reed-Relay OFF
B1=OFF
Again? no
>YES<
B1:Output Test
↕
Reed-Relay ON
B1=ON
Again? no
>YES<
Error-val. delay
>DAMPING<
edit
Error-val. delay
10
s