18 Outputs
150
UMFLUXUS_F7V4-6-2EN, 2017-10-01
18
Outputs
If the transmitter is equipped with outputs, they have to be installed and activated before they can be used:
• assign a measuring channel (source channel) to the output (if the transmitter has more than one measuring channel)
• assign the physical quantity (source item) to be transmitted to the output by the source channel, and the properties of the
signal
• define the behavior of the output in case no valid measured values are available
• activation of the installed output in the program branch
Output Options
18.1
Installation of an Output
All outputs are installed in
Special Funct.\SYSTEM settings\Proc. outputs
.
The source items and their scroll lists are shown in Tab. 18.1.
Note!
The configuration of an output will be stored at the end of the dialog. If the dialog is quit by pressing
key BRK, the changes will not be stored.
Select
Special Funct.\SYSTEM settings\Proc. outputs
. Press ENTER.
Select the output to be installed. Press ENTER.
The scroll list contains all available outputs. A tick
after a list item indicates that this out-
put has already been installed.
Select
active
if all current outputs to be installed are to be operated as active current out-
puts.
Select
passive
if all current outputs to be installed are to be operated as passive current
outputs.
This display will only be indicated if the transmitter is equipped with switchable current out-
puts.
This display will be indicated if the output has not been installed yet. Select
yes
. Press EN-
TER.
If the output has already been installed, select
no
to reconfigure it or
yes
to uninstall the
output and to return to the previous menu item to select another output. Press ENTER.
Select in the scroll list the measuring channel to be assigned to the output as the source
channel. Press ENTER.
This display will not be indicated if the transmitter has only one measuring channel.
Select the physical quantity (source item) to be transmitted from the source channel to the
output.
If a binary output is configured, only the list items
Limit
and
Impuls
will be displayed.
SYSTEM settings
↕
Proc. outputs
Install Output
↕
Current I1
(
)
Loop I1,I2
passive >ACTIVE<
I1 enable
no
>YES<
I1 disable
>NO<
yes
I1 Source chan.
↕
Channel A:
I1 Source item
↕
Measuring value