8 Start-up
FLUXUS F502TE
8.5 Language
69
UMFLUXUS_F502TEV1-5EN, 2020-05-29
8.5
Language
The language can be selected with the following HotCodes:
When the last digit has been entered, the main menu is displayed in the selected language.
The selected language remains activated when the transmitter is switched off and on again. After an initialization of the
transmitter, the language is reset to the default language.
8.6
Initialization
During an initialization (INIT) of the transmitter, the settings in the program branches
Parameter
and
Output Options
and some of the settings in the program branch
Special Funct.
are reset to the default settings of the manufacturer.
Proceed as follows to execute an initialization:
• When switching on the transmitter: keep BRK and C pressed.
• During the operation of the transmitter: press BRK, C and ENTER at the same time. A reset is executed. Release only
ENTER. Keep BRK and C pressed.
After the initialization has been executed, the message
INITIALISATION DONE
is displayed.
After the initialization, the remaining settings of the transmitter can be reset to the default settings and/or the stored measured
values can be deleted.
• Select
yes
to reset the remaining settings of the transmitter to factory default or
no
to keep them unchanged.
• Press ENTER.
If
yes
is selected, the message
FACTORY DEFAULT
DONE
will be displayed.
• Select
yes
to delete the stored measured values or
no
to keep them.
• Press ENTER.
This display will only be indicated if measured values are stored in the transmitter.
function
HotCode
reset of the display to medium contrast
555000
language
9090xx
manual input of the lower limit for the inner pipe diameter
071001
activation/deactivation of the SuperUser mode
071049
change of the transmission parameters for service interface RS232
232-0-
language
HotCode
Dutch
909031
French
909033
Spanish
909034
Czech
909042
English
909044
Polish
909048
German
909049
FACTORY DEFAULT
Delete Meas.Val.