17 SuperUser mode
FLUXUS F502TE
17.10 Display of the last valid measured value
127
UMFLUXUS_F502TEV1-5EN, 2020-05-29
17.10 Display of the last valid measured value
If the signal is not sufficient for a measurement,
UNDEF
is normally displayed. Instead of
UNDEF
, it is also possible to display
the last valid measured value.
• Select the menu item
Special Funct.\SYSTEM settings\Measuring\Miscellaneous
.
• Press ENTER until the menu item
Keep display val
is displayed.
• Select
on
to activate the display of the last valid measured value or
off
to deactivate it.
• Press ENTER.
17.11 Displays during the measurement
Besides the normal information (see section 9.4) the following parameters can be displayed during the measurement in
the SuperUser mode:
Special Funct.\SYSTEM settings\Measuring\Miscellaneous\Keep display val
display
explanation
t
transit time of the measuring signal in the fluid
c
sound speed
REYNOLD
Reynolds number
VARI A
standard deviation of the signal amplitude
VARI T
standard deviation of the transit time of the measuring signal
dt-norm
transit time difference standardized to the transducer frequency
fluid density