Example A
•
Target fill weight = 5 g
•
Actual fill weight measured by the balance = 5.1175 g
•
Upper auto recalibration tolerance limit = 7.5 g
•
Lower auto recalibration tolerance limit = 2.5 g
While pump is dispensing, a filled vial is placed on the balance. The measured
weight falls within the auto recalibration tolerances. The pump automatically enters
calibration mode and performs a calibration.
Example B
•
Target fill weight = 5 g
•
Actual fill weight measured by the balance = 2.2010 g
•
Upper auto recalibration tolerance limit = 7.5 g
•
Lower auto recalibration tolerance limit = 2.5 g
While the pump is dispensing, a foreign object is placed on the balance. The
measured weight falls outside the auto recalibration tolerances. The pump will
continue dispensing.
23.3.3 Pausing a batch
To pause a batch press
. Once the current fill has been completed, the
batch will pause.
When a batch is paused, the following options are available:
•
Resume manual batch
— This can be hidden, refer to "Operator access
•
Add note to report
—"Add note to report" on page 89
•
End batch
•
Batch size
•
Fluid sample
To resume filling press
.
PF7/PF7+ Peristaltic Filler with QC14 Pumphead User Manual