22.2 Single-fill calibration without a balance connected
1.
Place the container under the filling nozzle.
2.
Press
or use the external input to start the first calibration fill.
Calibration value entry
3.
Enter the dispensed value using the numeric keypad.
4.
Press
.
Calibration outside an acceptable level
If the calibration value entered is outside of ±95% of the target value:
•
A warning screen is displayed.
•
Calibration cannot continue.
•
Check the value was entered correctly.
•
Check the tube size is as per the recipe.
•
Check the fluid path is set up correctly.
Calibration with fill tolerance limits
Refer to "Fill tolerance" on page 63 to enable fill tolerance limits.
PF7/PF7+ Peristaltic Filler with QC14 Pumphead User Manual