20.18 Auto-recalibration tolerance
During manual filling
Auto-recalibration tolerance
defines an upper and lower
limit in relation to the target fill weight "Volume" on page 58
The pump will auto- recalibrate when the connected balance measures a weight
within these limits.
The limits should be set to allow an empty vial to be tared without recalibrating the
pump.
i
Manual batch only refer to "Dispense mode" on page 81.
Requires a connected balance.
i
If a correctly filled vial is outside of the auto-recalibration
tolerance, a manual recalibration can still be performed.
Refer to "Calibration whilst dispensing" on page 83.
20.19 Recalibration reminder
Sets the number of fills or time before a recalibration reminder. A message is
displayed and the pump beeps twice when reached.
i
Manual batch only Refer to "Dispense mode" on page 81.
20.20 Recalibration pause
Sets the number of fills or time before a recalibration pause. The batch is paused
and the pump beeps three times when reached. The batch cannot resume until a
recalibration is performed.
i
Manual batch only Refer to "Dispense mode" on page 81.
PF7/PF7+ Peristaltic Filler with QC14 Pumphead User Manual