25.14.2 Recalibration averaging
Recalibration averaging
recalibrates using a rolling average of the previous
stored recalibration values.
Value of one—no averaging.
Value of two or greater—incrementally adds recalibration value to rolling average.
i
Higher values
•
reduce negative effect of natural small variations in
fill volume
•
reduce positive effect of calibration if significant
difference between target fill weight and
recalibration value.
Optimum value depends on calibration frequency.
If the latest calibration value exceeds the fill tolerance limits ( "Fill tolerance" on
page 63), rolling average is reset and based only on new calibration value.
25.14.3 Weigh check option
On—Option
weigh check only
will display when calibrating during manual batch.
Refer to"Weigh check only" on page 84.
Off—the option
weigh check only
will not be displayed.
Weigh checking
Use
Weigh check only
to check the weight at any point during manual batch:
1.
Turn
Weigh check option
on
2.
Start a manual batch
3.
Do a calibration
4.
When the weight check only screen is displayed, select
weigh check only
i
The calibration value is saved in the batch report but the
calibration value used for dispensing remains unchanged.
25.14.4 Remove from batch option
On—Options to
remove from batch
or
include in batch
display
when calibrating
during manual batch. Refer to "Remove from batch option" on page 84.
Off—Every fill is included in batch.
25.15 Reporting
25.15.1 Batch reports
On—Batch report generated when new batch starts.
Off—No report generated.