![Flexicon PF7 User Manual Download Page 109](http://html1.mh-extra.com/html/flexicon/pf7/pf7_user-manual_2304659109.webp)
25.10 Session timeout period
Sets period of inactivity to automatically log out.
i
If FDA 21CFR Part 11 compliance is activated, this cannot
be disabled.
i
Log in will return the user to recipe mode. In event of
power failure if a batch was in progress the user is
returned to complete the batch.
25.10.1 Session timeout while pumping fluid
If the session timeout occurs while the pump is priming, calibrating or dispensing.
Pressing
will complete the current fill and then stop, or pressing
will stop the pump immediately.
The password is not required to stop the pump.
25.11 Password expiry duration
i
FDA 21CFR Part 11 compliance only. Refer to "Enable
United States Food and Drug Administration (FDA) 21CFR
Part 11 Compliance" on page 47.
Sets period in weeks before user required to set new password.
•
Maximum - 52 weeks
•
Minimum - 1 week
The pump cannot be operated until a new password is set. New passwords must be
different from the previous 5 passwords.
A setting of 0 weeks will ask the user to reset the password at every login. This
feature is for testing purposes.
PF7/PF7+ Peristaltic Filler with QC14 Pumphead User Manual
109
EN
EN
EN
EN
EN
EN
EN
EN
EN
EN
EN
EN
EN
EN
EN
EN
EN