9.1.1 Replacing the filters AP BIO 400, AP BIO 500 and AP BIO 1000
Replacing the pre-filter
Fig. 9-1
Replacing the pre-filter AP BIO 400, AP BIO 500 and AP BIO 1000
1: Pre-filter ISO ePM1 55% (F7)
2: Upper bottons
3: Safety lock
4: Access panel
•
Unlock the door (3) with a special access key at the door center.
•
Remove the access door (4) by pushing the two upper buttons (2).
•
Open the plastic bag, remove the new pre-filter and put it aside.
•
Slide out the used pre-filter very carefully, put it into the empty plastic bag of the new
filter and seal up with a tape.
•
Clean inner flat surfaces carefully with a cloth and disinfect with spray, observe the
exposure time of disinfection liquids of min. 30 sec. Also refer to “Cleaning the air
purifier” on page 27.
•
Slide in the new filter and make sure it is pushed it in all the way until it touches the
back wall of the unit.
Warning of contaminated filter surfaces!
When working on the unit or in case of damage to the components, there is a health
risk injury the contaminated filter surfaces.
•
Wear hygienic hand gloves and safety goggles
•
Wear protective respiratory mask
•
Exercise due caution and attention when carrying out this work.