6 Storage
Keep the accessories (contra-
angle, lip hook, file clip, touch
probe, protective silicon cover)
in sterilization packaging in a
dry and clean environment.
- Sterility cannot be guaranteed if packaging is
open, damaged, or wet.
- Check the packaging and the contra-angle
before use (packaging integrity, no humidity and
validity period).
7.
Maintenance
7.1
Calibration
Perform calibration after replacement or lubrification of the contra-angle (refer to Chapter 4.6).
7.2
Lubrification of the contra-angle
The contra-angle needs to be lubricated after cleaning and disinfection, but before sterilization.
1.
Screw the oil injection nozzle to the oil bottle (around 1 to 3 turns).
2.
Plug the nozzle into the end part of contra-angle
3.
Fill the contra-angle with oil until the oil flows out of contra-angle head part.
4.
Position contra-angle vertically for at least 30 minutes to let go the redundant oil under
gravity via the end part.
7.2.1
Warnings
a)
Use original oil injection nozzle only.
b)
Motor handpiece shall not be filled with oil.
7.3
Charging of the battery
1.
Insert the power adapter plug into the charging base power socket and make sure they
are correctly connected.
2.
Leave approximately 10cm around the charging base for easy access to inlet and the
power cord.
3.
Insert the motor handpiece into the charging base (the motor handpiece must be
correctly aligned with the charging base).
i.
While the motor handpiece is charging the LED indicator on the charging base flashes.
ii.
When the motor handpiece is fully charged, the LED indicator on the charging base
is always on.
4.
After charging, unplug the power adapter.
7.4
Replacement of the battery
1.
Power off the device.
2.
Use a tweezer or a screwdriver to open the rubber cover and then remove the screw.
3.
Remove the battery cover.
4.
Remove the old battery and disconnect the connector.
5.
Connect the new original battery and put it in the motor handpiece.
6.
Replace the cover and the screw.