background image

G

ENERAL

 

ADVICE

 

TO

 

TAKE

 

INTO

 

CONSIDERATION

 

FOR

 

REMOVAL

 

OF

 

FILLING

 

MATERIAL

» 

take 2 x-rays from different angles

» 

a precise appreciation of canal anatomy is very 

important

» 

fi nd entrance of canal and access the obturation 

material

» 

prepare, if needed, a cavity 1-2 mm (DR1) and 

drop solvent to soften obturation material. If 

necessary, use a heat plugger or ultrasonics.

C

ORONAL

 

THIRD

 

FILLING

 

REMOVAL

» 

gently engage DR1 rotating at 1000 rpm (1.5 Ncm) 

into the obturation material

» 

the active tip of DR1 facilitates the initial penetration

» 

don’t force the way, take out the fi le, clean it 

regularly and re-insert

» 

tip: clean and check the blade regularly

M

EDIAN

 

AND

 A

PICAL

 

THIRDS

 

FILLING

 

REMOVAL

» 

insert DR2 rotating at 600 rpm (0.7 Ncm)

» 

don’t force the way, clean and check the blade 

regularly

» 

continue to advance apically as long as fi lling 

material is visible on the fl ute

» 

if needed, drop solvent to help removal of 

obturation material

» 

establish fi nal WL and fi nalise shaping with 

standard NiTi RaCe fi le(s) 

M

EDIAN

 (2) 

TO

 A

PICAL

 (3) 

THIRD

C

ORONAL

 

THIRD

 (1)

DR1

@ 1000 rpm

DR2

@ 600 rpm

Instruction manual

Desobturation Kit

Active tip

Safety tip

D-RaCe fi les are intended to remove fi lling materials, 

such as gutta-percha, carriers, paste and resin-based 

materials, prior to reshaping procedures.