Page 21 of 34
SUPER SESI, by FIT
User manual, Revision 2.0
Figure 18. Inserting the silica capillary .
*The figure may differ from the latest design.
IMPORTANT NOTE
The back end of the silica capillary must be immersed in the
electrospray liquid
.
For optimum performance, use only the sharp silica capillaries
provided by FIT.
Operation
Adjusting the temperature
In order to control the temperatures, the SUPER SESI need to be powered, and the main switch
must be in the ON position.
Figure 19.
SUPER SESI main switch, IEC-C14 connector,
and temperature controllers.
The SUPER SESI incorporates two independent Proportional Integral Derivative (PID)
temperature controllers for the Sample Line and the Ionization Chamber. The electrospray
probe is heated by the ionization chamber by thermal conduction. In order to change the
temperature, follow the instructions below:
1.
In the corresponding temperature controller, set the new value using the Up and Down
Arrows (Figure 20). The new set point value is displayed at the right part of the controller.
The new set value is flickering until the enter key is hit.