14
Optical Connector
The mechanical life of the optical connector is about 500 mating cycles in a relatively clean environment.
To keep your EasyTest II in good operating condition, we have installed a short optical extension that act
as an optical Port Saver, the CFO-ET.
To protect your EasyTest II optical connector, always keep the CFO-ET connected to the main optical port
of your EasyTest II ‘’A’’ and connect the sensors to the Port Saver optical connector ‘’B’’.
You could also use it as an optical extension to test directly at the tank wall feedthrough.
When the optical connector of the CFO-ET worn out, simply order a new one by contacting
Main Optical
Connector ‘’A’’
Port saver optical
Connector ‘’B’’