Chapter 3 | Setup Menu
Press the “Enter” key to save the change.
3.2.8 Read Type
Set the Read Type as Continuous, Auto-Read or Timed to
define how measurements are performed and when
measurements are saved and exported.
Continuous:
In the Continuous measure mode, measurements are
continuously updated on the display and the stopwatch
(stabilizing) or checkmark (stable) icon indicates the
measurement stability status. Press the “Log/Export” key
to save a measurement to the data log and export.
Auto-Read:
In the Auto-Read measure mode, press the “Read” key to
start a measurement. When the measurement is stable, the
checkmark (stable) icon is shown and the measurement is
locked on the display until the “Read” key is pressed again.
The stable measurement is automatically saved to the data
log and exported.
Timed:
In the Timed measure mode, the measurement values are
continuously updated on the display. Measurements are
automatically saved to the data log and exported at the
selected time intervals, from 5 seconds to 60 minutes, the
entire time the meter is in the measurement mode.
Press the
or
key to scroll through the Setup
Menu list until the Read Type item is shown.
Press the
or
key to set the read type.
Press the “Enter” key to save the change.
3.2.9 Timed Interval
When Timed is set as the Read Type, set the time interval
from 5 seconds to 60 minutes. This time interval is used to
automatically save and export measurements.
Press the
or
key to scroll through the Setup
Menu list until the Timed Interval item is shown.
Press the
or
key to adjust the time value.
a.
Press the
key to increase the time value by
one second. Press and hold the
key to
quickly increase.
b.
Press the
key to decrease the time value by
one second. Press and hold the
key to
quickly decrease.
Press the “Enter” key to save the change.