Operator‘s Manual FERITSCOPE
®
FMP30
Page 89
Measuring
Measurement Capture Through Automatic Measurements
With automatic measurements, a specified number of readings is taken
with a selectable time interval between 2 measurements. In contrast to the
area measurement, all single readings will be stored (
This measuring mode can be advantageous, for example, when the
-ferrite
content distribution shall be determined along a line. If the probe is moved
by hand or a suitable mechanical device at a consistent speed along a line,
then the measurement points will be at equal distances.
Key sequ. /
Action
Detail of the display Explanation
Switch “Autom. Measurement” on from the
menu Service functions/Measurement/
Measuring Mode and select the desired
number of measurements and the time inter-
val of the measurements
(
10.7.4 ‘Measuring Mode - Standard/
Area Measurement/Automatic Measure-
ment’, beginning on Page 177).
Place the probe on the specimen. The dis-
plays the symbol
for as long as the
probe is on the specimen.
An audible signal sounds for every mea-
surement that occurred. All measurement
data are evaluated statistically.
Lift the probe off the specimen.
[Appl:]
: Application number
[Fe%]
: Displays
-
ferrite contents in ferrite
percent (measured variable of the open Ap-
plication 10.7.6
[Blck:]
: Number of the current block
[n=]
: Number of the measurements / block
size stored in the Application