Page 4
Operator‘s Manual FERITSCOPE
®
FMP30
Safety Information
1.6
Probe Handling
Fig. 1-1 Probe handling
During the measurement, the magnetic poles of the probes are placed di-
rectly onto the specimen. Observe the following to keep wear of the mag-
netic poles during the contacting measurement to a minimum:
1.7
Handling, Storage and Transport of the Base and the
Calibration Standards
The instrument is normalized and calibrated using a base and calibration
standards.
Proper condition of the calibration standards is an important prerequisite
To avoid breakage of the wiring, do not bend the probe connector
cable! The radius of rolled up probe connector cables should
always be at least 50 mm!
Place the probes speedily yet gently on the specimen surface!
Avoid hard impacts!
Do not drag the probe across the specimen surface.
Do not place standard probes on hot or acid-wetted surfaces; do
not immerse them in liquids. Special probe models are available
for such applications. You can obtain this data sheets from
FISCHER or from your authorized supplier.
Probe-
connector cable
R
50 mm !
R