Operator‘s Manual FERITSCOPE
®
FMP30
System Setup, Maintenance and Cleaning
3.3
Instrument: On/Off
3.3.1
Measurement Method of the Connected Probe
After powering up the instrument, the measurement method
[Ferrite]
ap-
pears on the display. The unit of measurement for displaying the readings
of the current Application is displayed next to it (
ning on Page 179 and 5.6.5 ‘Unit’, beginning on Page 52).
To avoid erroneous readings, no metallic objects must be in close
proximity to the probe tip when powering up the instrument.
The minimum distance is 25 mm.
Unit
Explanation
Fe % or
M%
-
ferrite content in ferrite percent or
-Martensit content
FN
-
ferrite content in ferrite numbers
If [Ferrite] flashes on the display, an Application has not yet been set up
using the connected probe. It is not possible to make measurements when
the display is flashing.
To make measurements, an Application must be set up using the con-
nected probe
(