Operator‘s Manual FERITSCOPE
®
FMP30
Page 169
Instrument Settings - Service Function Menu
10.6.3
Matrix Mode
Key sequ. /
Action
Detail of the display Explanation
Select
Device Mode
by pressing the arrow
keys
or
and confirm the selection with
ENTER
.
ENTER
Select the
Matrix mode
by pressing the ar-
row keys
or
and confirm the selection
with
ENTER
.
ENTER
This settings option appears only when “du-
al method” both is set for all Applications
set up with dual probes (
Select the desired mode by pressing the ar-
row keys
or
and confirm the selection
with
ENTER
.
[off]
: matrix mode is disabled, i.e., the
standard measuring mode is enabled
[on]
: Matrix measuring mode is enabled
[10/20/150]: Example for 10 Applica-
tions each with 20 blocks of 150 measure-
ments each.
Use
DEL
at any time to cancel the setting
procedure.