
Page 152
Operator‘s Manual FERITSCOPE
®
FMP30
Instrument Settings - Service Function Menu
10.2.8
Re-Initialization of the Instrument
The restoration of the default settings of the instrument is called re-initializa-
tion.
With a re-initialization of the instrument,
all Applications are deleted, i.e., the stored measurement data as well
as the normalizations and the corrective calibrations of all Applications
(thus, after a re-initialization, all Applications must be set up anew!),
all readings are deleted,
all settings in the service functions are reset to the default settings (i.e.,
to the factory settings) (with the exception of time, date and language);
thus, if necessary the settings will have to be made again,
all settings in the service functions are reset to the default settings (i.e.,
to the factory settings) (exception: language); thus, if necessary the set-
tings will have to be made again,
the coefficients of the master characteristic that is stored in the EEPROM
of the probe plug are not changed because the re-initialization con-
cerns only the memory of the instrument.
Key sequ. /
Action
Detail of the display Explanation
Select
System
from the service menu.
Use
ENTER
to confirm the selection.
ENTER
Select the
Initialization
by pressing the ar-
row keys
or
and confirm the selection
with
ENTER
.
ENTER
[Yes: DEL]
: Press
DEL
to carry out the re-
initialization.
[No: ENTER]
: Press
ENTER
to not carry
out the re-initialization.