7
Before operation, it is necessary to become familiar with
the freezer controls located on the header panel. A layout
of the controls is given below.
Begin operation by inserting the key into the key switch
located on the header panel. The key switch is packed
inside the envelope, which is shipped in the refrigerator
chamber. Turning the key switch to the ON ( | ) position
will energize the digital panel display.
Control Layout
The digital temperature controller is located on the header
panel (See Figure 3). When the unit is initially turned on,
the display will indicate current chamber temperature.
The temperature units will be indicated by the °C or °F
LED located just to the right of the temperature display.
The refrigerator is factory set at -15°C.
When the unit is first energized, there is a delay time
before the compressor will energize. When the delay
time is over, the Cool LED will illuminate, the compressor
will run and the chamber will cool
Setting the Temperature Control Setpoint
To change the set temperature, press and release the
Menu keypad once. The display will flash SP and the
Mode LED will be illuminated. The last set temperature
Operation
I
O
POWER
Units LED'S
Decrease Keypad
Cool LED
Key Switch
Temperature Display
Mode LED
Menu Keypad
Increase Keypad
Warning
This product is NOT approved for stor-
age of flammable or explosive materi-
als. Also, it is NOT approved for use
in hazardous locations containing
explosive atmospheres.
Figure 3