Shelves

Shipped inside each cabinet are 3 epoxy coated shelves

per door opening. Epoxy coated filler shelves are provid-

ed for units with more than one door. Shelf supports,

located in the plastic bag, are pushed into the side

pilasters, one in each corner, at the desired shelf height.

Replacement shelves are available individually. See

Replacement Parts for shelf part numbers. See Figure 2

for shelf installation.

Electrical Connection

Determine the total amount of current presently being

used by other apparatus connected to the circuit that will

be used by this freezer. It is critical that this added cur-

rent demand and other equipment on this circuit not

exceed the rating of the fuse or circuit breaker in use.

The frequency and nominal voltage requirements for the

unit are specified on the data plate, which is located on

the interiorʼs upper left side. Plug the unit only into a

power source that meets these requirements. Low line

voltage is often the cause of service complaints. With the

unit running, check that the line voltage is within ±10% of

that specified on the data plate.

The customer should have the wall receptacle and circuit

checked by a qualified electrician to verify the receptacle

is properly grounded and is connected to the proper

power supply. The following list gives the NEMA

plug/receptacle configuration for the corresponding freez-

ers.

6

U

NPACKING AND

I

NSTALLATION

Caution

If the unit was tilted in excess of 30

degrees, level it and do not apply

electrical power for a minimum of 12

hours.

Warning

Be sure that the power supply is the

same voltage that is specified on the

freezerʼs data plate.

Warning

For personal safety this unit must be

properly grounded. DO NOT under

any circumstance cut or remove the

ground prong from the power cord.

DO NOT use a plug adapter that elimi-

nates the ground prong.

Lab Freezers (115V)

Catalog Noʼs

NEMA Plug

NEMA Receptacle

13-986-142

5-20P

5-20R

13-986-144

5-20P

5-20R

Lab Freezers (230V)

Catalog Noʼs

NEMA Plug

NEMA Receptacle

13-986-143

L14-20

L14-20

13-986-147

L14-20

L14-20

Caution

Do not use an extension cord. Use of

an ungrounded cord or an overloaded

circuit VOIDS the compressor warran-

ty.

Figure 2

Caution

Catalog numbers ending in HK are

international 3-Wire systems and

should not be confused with 230V

domestic units which are 4-Wire sys-

tems. Consult a qualified electrician

before powering this unit.

Summary of Contents for Isotemp Plus 13-986-142

Page 1: ...Isotemp Plus General Purpose Laboratory Freezers Catalog Numbers 13 986 142 142R 13 986 143 143R 13 986 144 144R 13 986 147 147R 104215 Rev A...

Page 2: ...Casters 5 Shelves 6 Electrical Connection 6 Operation 7 Control Layout 7 Setting the Temperature Control Setpoint 7 Units Select 8 Calibration Offset 8 Hold Off Time 9 Temperature Digression 9 Automat...

Page 3: ...s intended application alterations or modifications will void the Warranty This product is not intended nor can it be used as a ster ile or patient connected device In addition this appara tus is not...

Page 4: ...such external evidence of loss or damage may result in the carrier s refusing to honor your damage claim The carri er will supply the form required to file such a claim Concealed Loss or Damage Conce...

Page 5: ...it After installation confirm that the unit was installed per these instructions that the required information was completed and that the form was faxed to the specified number Installing and Leveling...

Page 6: ...e customer should have the wall receptacle and circuit checked by a qualified electrician to verify the receptacle is properly grounded and is connected to the proper power supply The following list g...

Page 7: ...icated by the C or F LED located just to the right of the temperature display The refrigerator is factory set at 15 C When the unit is first energized there is a delay time before the compressor will...

Page 8: ...nd control point to a ref erence standard To set the temperature offset press and hold the Menu keypad for 5 seconds The display will flash oS offset followed by the last temperature off set value The...

Page 9: ...sing the Increase keypad the display will show the warmest temperature recorded since the unit was ener gized or the digression values were reset Pressing the Decrease keypad will display the coldest...

Page 10: ...t and lukewarm water or a solu tion of bicarbonate of soda 1 tablespoon per gallon of water is recommended for cleaning the interior and exte rior of the cabinet All surfaces should be rinsed and thor...

Page 11: ...ondensate evaporator pan located behind and below the condenser fan must be cleaned periodically to prevent foul odors and to operate efficiently Vacuum clean if dry or sponge clean with soap and wate...

Page 12: ...ot A clicking sound occurring about every 30 seconds indicates this protector is working Insufficient cooling Set temp is too high Reduce temperature setting verify Cool LED not lit Cool LED is on Con...

Page 13: ...t 1 800 438 4851 Item Part No Shelves SPN13986210E Compressor Relay K1 SPN88616 Temperature Controller SPN104161 International Version Controller SPN105402 Chart Recorder SPN104114 Chart Recorder Inst...

Page 14: ...K DEFROST SWITCH 1 BLK BLK YEL BLU L1 Compressor Relay K1 COMPRESSOR MOTOR CONDENSER FAN S K1 BRN DEFROST TIMER CUT OUT COIL ORN RED DEFROST LIMIT CONTROL WHT DEFROST HEATER CONDENSATE DRAIN HEATER OR...

Page 15: ...sor Relay Cool Relay R1 CONDENSATE DRAIN HEATER COMPRESSOR MOTOR CONDENSER FAN S DEFROST HEATER WHT BRN PNK R1 DEFROST TIMER CUT OUT COIL DEFROST SWITCH 1 ORN ORN DEFROST LIMIT CONTROL RED BLK YEL YEL...

Page 16: ...m the respective manufacturer whatever adjustment is reasonable in the light of the manufacturer s own warranty b The Company shall be released from all obligations under its warranty in the event rep...

Reviews: