Display
The Fisher Scientific accumet AE150 pH benchtop meter display has mode indicators for
pH, mV, relative mV (R.mV) and temperature readings.
The primary (upper) display shows the measured values of pH, mV or R.mV and the
secondary (lower) display shows the temperature reading in Celsius or Fahrenheit.
Special indicators such as units of measure, error message, icons and modes are arranged
around the measurements to give a simplified, uncluttered display.
Icon
Description
SETUP The meter is in the setup mode
MEAS
The meter is in the measure mode
CAL
The meter is in the calibration mode
MEM
Data is stored to the memory
READY The measured value is stable
HOLD
The displayed value has been frozen
Prompts user to select buffer during calibration; or indicates buffer error when flashing with
ERR icon
Indicates electrode error when flashing with ERR icon
ERR
An error in operation has occurred
ATC
Indicates Automatic Temperature Compensation is activated; when ATC is not displayed,
temperature probe is not connected and manual temperature compensation is activated
Units of Measure
Indicates buffer
values of USA or
NIST buffer set