FISCHER Mess- und Regeltechnik GmbH
Operation | 5
BA_EN_DE90
61
5.5.1.4 Number format C1
Directory: \Configuration\Channel 1\Number format C1
Level: 3
\...\...\Number format C1
1
2 3
4
±123456
±12345.6
±1234.56
±123.456
S
R
U
R
R
Fig. 81:
Number format C1
The number of decimal places can be determined with this menu. All theoretic-
ally possible variants are made available for selection.
The decimal places are limited by the measuring range. There are 8 characters
available with signs, decimal points and number value. The measuring data dis-
play can have less decimal points than set in the number format.
For example:
Set number format:
±123.456
Current measuring value:
-1234.567
Displayed measuring value:
-1234.57
Only two decimal points are shown, as otherwise the maximum number of 8
characters would be exceeded. The last decimal place is rounded.