26
FISCHERSCOPE
®
X-RAY
Set up
3.5
Switching on the Instrument
Before you begin
The instrument was placed correctly.
The shipping lock was removed.
All connections were established.
Make sure that the FISIM (Fischer Software Identification Module) is plugged into a USB port
of the computer.
Make sure that the high voltage key switch is in the on-position. For that purpose open the cover
of the measurement chamber. The key switch is located at the detector unit of the instrument.
To switch on the instrument and the PC
1. Switch on the instrument by using the main switch on the right side of the instrument.
2. Switch on the PC and the monitor.
Windows will be started.
3. To start WinFTM, double click the WinFTM icon
.
4. In the WinFTM start window click
OK
.
The software will be started and the communication between the instrument and the software
will be initialized. The product that had been selected prior to powering off will be called auto-
matically.
What you can do next
You can now perform measurements.
The following activities must not be performed immediately after power up, but only after
about half an hour of measuring time:
Calibration/Normalization
Monitoring measurement devices including corrective measures
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......