212
FISCHERSCOPE
®
X-RAY
Features of WinFTM BASIC and LIGHT
2.8
Optical Functions
Function
Description
Video image of the speci-
men
The specimen is shown in the video window inside the WinFTM
main window.
Scaled crosshairs
Scaled crosshairs are electronically inserted into the video image.
They show the real size and position of the measurement spot on
the surface of the specimen. The color of the crosshairs is selecta-
ble.
Auto-focus (on models with
motor-driven z-axis)
The detector unit moves up and down automatically. The auto-fo-
cus function allows a significantly more precise setting of the meas-
urement distance than subjective manual focusing. Thus, the
reproducibility of the measurement result is increased.
Measurement of the dis-
tance between the test spot
and the detector / DCM
method
Focusing the video image provides the correct measuring distance.
WinFTM takes into account these distance values, when comput-
ing the coating thickness measurement data, using the DCM Meth-
od (Distance Controlled Measurement).
This allows for automated testing of complex surface geometries
and measurements in recesses.
Displaying programmed
points in the video image
(on models with motor-driv-
en XY-stage)
You can display programmed points in the video image.
Software-based video
zoom function
You can zoom the video image in steps of 100, 200, 300 and
400 percent. Additionally, you can display the video image in full
screen mode.
Saving video images to-
gether with the measured
values
You can edit the video image with a graphic program of your
choice e.g. for adding comments or dimensions. When you save
the video image, it will be assigned to your measured values. Ad-
ditional text notes will be stored, linked to the video image.
Recognition of vertical or
horizontal strips (on models
with motor-driven XY-stage)
The instrument can recognize strip structures on specimens and re-
adjust the XY position along this structure. Thus, you can make sure
that on specimens with geometric tolerances, the measurement is
always performed at the correct position.
Image/Pattern Recognition
(on models with motor-driv-
en XY-stage and WinFTM
extension SUPER activated)
The instrument can recognize a predefined structure and readjust
the XY position of the XY-stage along this structure automatically.
Thus, you can compensate deviations in positioning during meas-
urements on more specimen (e. g. for routine tests in production).
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......