Pre-run for determining the control limits
Measurement device monitoring for the Fischerscope X-RAY
FISCHERSCOPE
®
X-RAY
119
With a Fischerscope XRAY with WinFTM
software, the spectra of
the ”pure elements” are not measured anew during the
normalization but instead are accepted from the spectra library that
contains their spectra when the switch ”Measure elements” of the
DefMA is disabled. In this manner, the measurement application is
not changed during normalization, and therefore not corrected,
except for a potential substrate material correction. For such
measurement applications, normalization cannot provide any new
information.
The measurement application should be checked once more after
every corrective measure. If the limits are still violated, we
recommend a re-calibration of the measurement application. Due to
its high stability, corrections of the measurement applications of a
Fischerscope XRAY are typically required only after several weeks
or months.
NOTE:
Note:
All measurements for monitoring of the mea-
surement devices as well as all potential corrections
should only be performed once the instrument has
been measuring for a sufficient length of time, that is,
it has generated ”radiation”. 15 – 30 min of operating
time are sufficient for this purpose. Thus, measure-
ment device monitoring is not recommended directly
after power-up of the instrument, even though regular
measurements can be made.
15.9 Pre-run for determining the control limits
NOTE:
Note
: The equations and factors used here are all de-
scribed in the equation collection of the operator’s
manual for the Fischerscope XRAY. The control limits
can be calculated using the ”calculator” functions in
the control chart of WinFTM. Beyond that a spread-
sheet that contains the equations beneath for the cal-
culation of the control limits can be used for the pre-
run.
The pre-run consist of a minimum of m = 10 groups each with
n = 5 single readings, i.e., a total of 50 measurements. The
measurement application and the identity designations must be
documented.
The single readings for the pre-run should be carried out
successively using the setting parameters (measurement time,
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......