FISCHERSCOPE
®
X-RAY
37
WinFTM File Structure
6
WinFTM File Structure
To ensure correct measurements, a set of specific files needs to be set up. The process
flow for setting up the files needed for measurements, is the same on all FISCHERSCOPE
X-RAY instruments:
Fig. 6-1: Process flow for preparing measurements
The essential content of the files product, measurement application and Def.MA and their
relations among each other is shown in the following figure:
Fig. 6-2: Content of the files Def.MA, measurement application and product
Def.MA:
Definition of the measurement
settings of the application
Product
Measurement results and their presentation
Image of the product
Measurement time and number of measurements
Export masks
Measurement Application
Calibration data (Calibration standard set)
Measurement time during calibration
Number of measurements during calibration
Anode current, aperture
Def.MA
Measurement mode
Type and order of the coatings
Tube voltage
Type and composition of the calibration standards
Treatment of interfering spectra
Special evaluation methods
Summary of Contents for FISCHERSCOPE X-RAY 5000 Series
Page 22: ...22 FISCHERSCOPE X RAY Performing Manual Measurements Deleting Measurement Readings...
Page 36: ...36 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Page 40: ...40 FISCHERSCOPE X RAY WinFTM File Structure Product...
Page 118: ...118 FISCHERSCOPE X RAY Def MA Display the Measurement Mode...
Page 124: ...124 FISCHERSCOPE X RAY Calibration...
Page 142: ...142 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Page 156: ...156 FISCHERSCOPE X RAY Addendum Assignment of the Electrical Connections...
Page 183: ...WinFTM 183...