FISCHERSCOPE
®
X-RAY
27
Statistics Field
User Interface of the WinFTM Software
5.3
Statistics Field
The statistical results of the last block are displayed in this field (see Fig. 5-1 on Page 23).
A block is a test series. The test series has a first and a last value. The last value marks
the end of the block. An additional measurement value belongs to the next test series.
The following features of a block are displayed in the statistics field: Number of the block,
Mean value, Standard deviation, Coefficient of variation, Number of measurements, Order
no. and Operator.
5.4
Measurement Value Field
This field either displays the last measurement value or a list of all measurement values
acquired with the current product.
You can choose between three different display modes:
Large number presentation
List presentation
SPC chart presentation (Statistical Process Control Chart)
5.4.1
Large Number Presentation
In this display mode the current measurement value (or the last values, if several charac-
teristics are measured) is displayed in large digits. In addition to the measurement value,
the measurement uncertainty is displayed as well. The measurement uncertainty depends
on the measurement conditions (measurement time, measurement distance, measure-
ment application).
If you have set up tolerance limits, the measurement value is displayed with colored back-
ground. If the measurement value is within the tolerance limits, the measurement value is
displayed on green background, otherwise on red background.
5.4.2
List Presentation
The list of measurement values are presented in the order of the measurements. In addi-
tion to the measurement values, the date and time, the name of the operator and the toler-
ance limits are displayed as well. The name of the operator and the tolerance limits are
displayed, if they were previously entered.
Summary of Contents for FISCHERSCOPE X-RAY 5000 Series
Page 22: ...22 FISCHERSCOPE X RAY Performing Manual Measurements Deleting Measurement Readings...
Page 36: ...36 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Page 40: ...40 FISCHERSCOPE X RAY WinFTM File Structure Product...
Page 118: ...118 FISCHERSCOPE X RAY Def MA Display the Measurement Mode...
Page 124: ...124 FISCHERSCOPE X RAY Calibration...
Page 142: ...142 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Page 156: ...156 FISCHERSCOPE X RAY Addendum Assignment of the Electrical Connections...
Page 183: ...WinFTM 183...