16
FISCHERSCOPE
®
X-RAY
Switching the Instrument on and off
Switching off the Instrument
6. In the WinFTM start window click
OK
.
The software will open in the ready-to-measure mode (Communication on). The
product that had been selected prior to powering off will be called automatically.
What you can do next
You can now perform measurements.
3.2
Switching off the Instrument
To switch off the instrument and the PC
1. Switch off the instrument by using the Off/On rocker switch at the rear of the instru-
ment.
2. To terminate WinFTM, select
File > Exit
.
3. To shut down the PC, select
Start > Shut down
.
4. Switch off the monitor and the printer, if present.
The following activities must not be performed immediately after power up, but
only after about half an hour of measuring time:
Calibration and normalization
Monitoring measurement devices including corrective measures
*
)
(
'
&
%
Off/On rocker switch